Microlinks UM05 - user manual Manuale Utente Pagina 61

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 72
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 60
Microscope Application Program http://www.ViTiny.com
61
Copyright © 2008-2012 MicroLinks Technology Corp.
Fig. 5-60 Input the actual measurement length
The calibration frame can be freezed by pressing the right button on the
mouse. The freezed frame helps for calibrating in a stable condition. Put the
calibrator under the machine and adjust focus to clearly see the scale. Key in “400”
length in the check-box of “Input Measurement Length”. Use mouse to draw a
yellow line for 4.00mm.Click “Finished” to complete and save the new calibration
figure.
The new setting scale (Calibration Scale [New]) show on the frame. See
Fig.5-61.
Fig. 5-61 Draw a line in actual length
5.4.6Scale information setting “
This setup screen indicates to set whether to show the unit value or not, see Fig. 5-60.
Choose “length” means to display the length of the line, by capital letter “L” expression, if
length is 2.33mm, it will show L=2.33mm besides the line. If choose none”, no value will
show on the dialogue. If show “P” means “perimeter”, “A” means “Area”, “C” means
“Circumference, and “R” means ”radius”
Vedere la pagina 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 71 72

Commenti su questo manuale

Nessun commento